Evaluation of the Dade Behring Dimension RxL Clinical Chemistry Analyzer by Smiljana Čuka, Štefica Dvornik, Kristina Draženović, Jasna Mihić

The performance of the Dade Behring Dimension RxL clinical chemistry analyzer was evaluated according to the guidelines of the European Committee for Clinical Laboratory Standards. The following analytes were tested: glucose, urea, creatinine, albumin, phosphorus, cholesterol, triglyceride, uric acid, magnesium, sodium, potassium, chloride, calcium, iron and total bilirubin. The Dade Behring Dimension RxL was compared with the Hitachi 704, Bayer RA-1000, Ektachem 250 and Chiron 865 depending on available tests on these analyzers. Coefficients of correlation showed high correlation between compared analyzers. Other performances (intra- and inter-assay variation, carry-over and interferences) of the analyzer were satisfactory.