Background: A rapid and reliable screening test for urinary tract infection (UTI) is needed to reduce the turn-around time and to rule out negative results of urine culture. The aim of this study was to evaluate the performance of BACT count and BACT-Info flag of the UF-5000 for screening for UTI.
Methods: A total of 1,063 urine specimens from April to September 2019 were included in this study. We evaluated the diagnostic performance of white blood cell (WBC) count, BACT count, BACT-Info flag, and UTI flag in UF-5000 by comparing with the urine culture results.
Results: Of the urine specimens, 16.7% were culture-positive (≥ 105 CFU/mL) with 15 being yeast positive. A BACT count of > 685.3/µL showed the best diagnostic performance with 93.8% sensitivity and 90.2% specificity. We confirmed that the combination of BACT count (685.3/µL) and BACT-Info flag would be appropriate to use in a clinical laboratory (sensitivity 91.5%, specificity 90.5%). Based on this combination, the sensitivity and specificity of the Gram-negative flag were 95.5% and 94.8%.
Conclusions: We recommend the use of a combination of BACT count (685.3/µL) and BACT-Info for UTI diagnosis. This combination is more appropriate for Gram-negative bacteria, and it would be useful for selecting empirical treatment.