Multicenter Evaluation of a New High-Throughput HbA1c Testing Platform by R. Imdahl, R. Roddiger, E. Casis-Saenz

Background: This non-interventional, multicenter study with anonymized leftover patient samples was performed to evaluate the reliability and analytical performance of the novel high-throughput HbA1c cobas c 513 analyzer.
Methods: A performance evaluation was carried out at three sites to validate the overall system functionality, user interaction and analytical performance of the new cobas c 513 analyzer using the Tina-quant® HbA1c Gen. 3 assay.
Results: HbA1c applications for both whole blood and hemolysate samples show a high precision using both quality control materials and pools of whole blood or hemolysates. The method comparison of HbA1c Gen. 3 on the cobas c 513 with HbA1c Gen. 2 on the Menarini HA-8180V using 249 whole blood samples shows high concordance. Moreover, analyte concentrations as measured by the cobas c 513 and Tosoh G8 and HbA1c Gen. 2 on COBAS INTEGRA® 800 CTS are comparable.
Conclusions: The cobas c 513 has proven to be a reliable system with excellent analytical performance of the Tinaquant® HbA1c Gen. 3 assay in high throughput laboratories.

DOI: 10.7754/Clin.Lab.2016.160633