Background: To build and evaluate intrinsic coagulation factor assays which can eliminate the effect of lupus anticoagulants (LAC). Methods: Commercial silica clotting time confirmatory (SCT-C) reagent containing sufficient synthetic phospholipid and routine activated partial thromboplastin time (APTT) reagent were each used for one-stage detection of FVIII, FIX, and FXI activities, in samples with or without LAC, and the results were compared. Results: For samples without LAC, consistent results of FVIII, FIX, and FXI using both SCT-C reagent and APTT reagent were obtained. For samples with LAC, the assays with SCT-C reagent not only could eliminate the effect of strong lupus anticoagulants but also needed fewer dilutions than that with routine APTT reagent. Conclusions: The intrinsic factor detections by SCT-C reagent are credible and convenient to be used for samples with LAC.
DOI: 10.7754/Clin.Lab.2015.151106
|