Performance Evaluation of a Novel Chemiluminescence Assay Detecting Treponema Pallidum Antibody as a Syphilis Screening Method by Qixia Chen, Jingna An, Chenli Rao, Tingting Wang, Dongdong Li, Shu Feng, Chuanmin Tao

Background: Syphilis is a major concern to global public health with increasing incidence. So its screening test should have sufficient sensitivity and specificity. We evaluated the performance of the Lumipulse® G TP-N assay detection for syphilis screening and compared it with the InTec ELISA test kit for TP, which is widely used.
Methods: Samples of several patient groups including 133 clinical and serologically characterized syphilitic sera, 175 samples containing potentially interfering agents, and 2290 unselected samples submitted for routine screening were detected by both the Lumipulse® G TP-N assay and the InTec ELISA test kit for TP. Inconsistent samples were confirmed by RecomLine Treponema IgG, IgM immunoblot.
Results: Coefficient of variations of the Lumipulse® G TP-N assay at both levels were below 5% and of the InTec ELISA test kit for TP both over 5%. The sensitivity of the Lumipulse® G TP-N assay and the InTec ELISA test kit for TP were 100% for all stages of syphilis. The two methods had consistent analytical specificity of 100% (95% CI: 97.21 - 100.00), while the clinical specificity was 100% (95% CI: 99.79 - 100.00) and 99.82% (95% CI: 99.51 - 99.94), respectively. Between them, Spearman’s correlation coefficient was 0.455 and kappa value was 0.986. The overall sensitivity and specificity of the Lumipulse® G TP-N assay was higher than the InTec ELISA test kit for TP (sensitivity: 100.0 versus 99.5, specificity: 100.0 versus 99.8).
Conclusions: The automated Lumipulse® G TP-N assay demonstrated excellent diagnostic sensitivity and specificity when evaluated as a screening test for syphilis. Thus, it can be an alternative to the treponemal screening test.

DOI: 10.7754/Clin.Lab.2015.150625