|
Background: To evaluate the performance of Sysmex XE-5000 analyser for detecting nucleated red blood cells (NRBCs) in peripheral blood and investigate the clinical application of this analyser. Methods: The absolute NRBC counts (NRBC#) and percentage (NRBC%) of 137 blood specimens (NRBC-positive according to the DIFF channel of the analyser) were determined in the NRBC channel of the analyser. The intraassay imprecision, carryover rate, and linear range of the analyser were evaluated. The NRBC% of the blood sample was detected with a microscope, and the difference between two methods was analysed. Results: The intra-assay imprecision of the analyser for detecting NRBC# in specimens with high, moderate, and low Q-flag values were 2.10%, 3.26%, and 11.62%, respectively, and the imprecision for detecting NRBC% were 3.79%, 5.80%, and 13.33%, respectively. The carryover rates of the analyser for detecting NRBC# and NRBC% were 0.51% and 0.26%, respectively. Conclusions: Sysmex XE-5000 analyser had good linearity in NRBC# (i.e., 0 /L to 18 × 109 /L). The NRBC%s of the two methods did not significantly differ (p = 0.716). The analyser can completely replace the traditional microscope for clinically classifying and counting NRBCs.
DOI: 10.7754/Clin.Lab.2013.130412
|