Participation and Performance in INSTAND Multi-Analyte Molecular Genetics External Quality Assessment Schemes from 2006 to 2012 by Friedrich E. Maly, Roman Fried, Michael Spannagl

Background: INSTAND e.V. has provided Molecular Genetics Multi-Analyte EQA schemes since 2006.
Methods: EQA participation and performance were assessed from 2006 - 2012.
Results: From 2006 to 2012, the number of analytes in the Multi-Analyte EQA schemes rose from 17 to 53. Total number of results returned rose from 168 in January 2006 to 824 in August 2012. The overall error rate was 1.40 ± 0.84% (mean ± SD, N = 24 EQA dates). From 2006 to 2012, no analyte was reported 100% correctly. Individual participant performance was analysed for one common analyte, Lactase (LCT) T-13910C. From 2006 to 2012, 114 laboratories participated in this EQA. Of these, 10 laboratories (8.8%) reported at least one wrong result during the whole observation period. All laboratories reported correct results after their failure incident.
Conclusions: In spite of the low overall error rate, EQA will continue to be important for Molecular Genetics.

DOI: 10.7754/Clin.Lab.2013.130411