Background: A number of automated devices for pretransfusion testing have recently become available. This study evaluated the Immucor Galileo System, a fully automated device based on the microplate hemagglutination technique for ABO/Rh (D) determinations. Methods: Routine ABO/Rh typing tests were performed on 13,045 samples using the Immucor automated instruments. Manual tube method was used to resolve ABO forward and reverse grouping discrepancies. D-negative test results were investigated and confirmed manually by the indirect antiglobulin test (IAT). Results: The system rejected 70 tests for sample inadequacy. 87 samples were read as “No-type-determined” due to forward and reverse grouping discrepancies. 25 tests gave these results because of sample hemolysis. After further tests, we found 34 tests were caused by weakened RBC antibodies, 5 tests were attributable to weak A and/or B antigens, 4 tests were due to mixed-field reactions, and 8 tests had high titer cold agglutinin with blood qualifications which react only at temperatures below 34°C. In the remaining 11 cases, irregular RBC antibodies were identified in 9 samples (seven anti-M and two anti-P) and two subgroups were identified in 2 samples (one A1 and one A2) by a reference laboratory. As for D typing, 2 weak D+ samples missed by automated systems gave negative results, but weak-positive reactions were observed in the IAT. Conclusions: The Immucor Galileo System is reliable and suited for ABO and D blood groups, some reasons may cause a discrepancy in ABO/D typing using a fully automated system. It is suggested that standardization of sample collection may improve the performance of the fully automated system.
DOI: 10.7754/Clin.Lab.2013.121011
|