You have to be registered and logged in for purchasing articles.

Abstract

Candidate Reference Measurement Procedures for Chloride, Potassium, Sodium, Calcium, Magnesium, and Lithium by Inductively Coupled Plasma (Isotope Dilution) Sector Field Mass Spectrometry (ICP-(ID) SFMS) in Serum by U. Kramer, M. Kress, H. Reinauer, M. Spannagl, P. Kaiser

Background: Standardization of the measurement of electrolyte concentrations in se-rum is of considerable interest for quality assurance in patient care. To promote the ongoing process of standardization we developed candidate reference measurement procedures of highest metrological order for Cl, K, Na, Ca, Mg, and Li using ICP-(ID) SFMS.
Methods: Serum samples were diluted with 4 mmol/L nitric acid and were spiked with the internal standard for quantification, separately for each analyte. The samples were introduced in the ICP-SFMS device by continuous infusion using a peristaltic pump. The measurement results were compared with reference measurement procedure values obtained by atom absorption spectroscopy, flame emission spectroscopy, and coulometry. The measurement accuracy and precision was calculated by analyzing cer-tified reference materials and EQAS samples.
Results: The mean coefficient of variation (CV) of the ICP-MS procedures for the se-rum samples was 0.65% for Cl, 0.46% for K, 0.51% for Na, 0.77% for Ca, 0.78% for Mg, and 0.58% for Li. The mean bias from target values of NIST certified reference materials was +0.85% for Cl, -0.46% for K, +0.68% for Na, -0.21% for Ca, +0.27% for Mg, and -0.39% for Li.
Conclusions: Candidate reference measurement procedures for 6 electrolytes were de-veloped by high performance magnetic sector field ICP-MS fulfilling the require-ments of ISO 15193:2009 for reference measurement procedures with traceability to SI according to ISO 17511:2003 and can be used for setting target values in EQAS and for certification of reference materials.

DOI: 10.7754/Clin.Lab.2012.120902