You have to be registered and logged in for purchasing articles.

Abstract

Comparison of Architect I 2000 for Determination of Scc with IMX Determination of SCC with Different Methods by Nafija Serdarevic, Adina Elena Stanciu

Background: SCC (squamous cell carcinoma) antigen (SSCA), a tumor marker, is already used for the diagnosis and follow-up of carcinoma of the cervix and lungs. The aim of our study was the determination of SCCA concentration in 96 serum samples using CMIA (chemiluminesecent microparticle immnoassay) on an Architect i2000 (Abbott Diagnostics) and MEIA (microparticle enzyme immnoassay) on an IMx Chemistry Analyzer (Abbott Diagnostics).
Methods: In our investigation we used a CMIA (chemiluminesecent microparticle immnoassay) Architect i2000 and MEIA (microparticle enzyme immnoassay) Imx Chemistry Analyzer for SCCA determination.
Results: All patients were hospitalized at the Department of Gynecologic Oncology and Department for Oncology at the University Clinics Center of Sarajevo. The normal serum range of SCCA should be < 0.1 μg/L. The quality control, precision and accuracy of the Architect i2000 were assessed. The quality control was done using quality control serums for low (X̅ = 2.17 μg/L), medium (X̅ = 10.23 μg/L), and high (X̅ = 48.99 μg/L) range. The precision for the Architect i2000 has a CV of 3.44% to 4.44%. We established that the main difference between Architect i2000 and IMx was statistically significant at p < 0.05 according to Student's t-test. The correlation coefficient was r = 0.990. The some samples have higher concentrations on the Architect than on the IMx because the sensitivity of the Architect assay is high (< 0.1 μg/L).
Conclusions: The CMIA Architect technology is an applicable method significant in diagnostic of SCCA.

DOI: 10.7754/Clin.Lab.2013.121133